Supplementary Material for: Intraocular Endoscopy Resolved Tube Occlusion of an Ahmed Glaucoma Valve
收藏Figshare2021-08-24 更新2026-04-28 收录
下载链接:
https://figshare.com/articles/dataset/Supplementary_Material_for_Intraocular_Endoscopy_Resolved_Tube_Occlusion_of_an_Ahmed_Glaucoma_Valve/16419018
下载链接
链接失效反馈官方服务:
资源简介:
We report a case in which intraocular endoscopy clarified the cause of Ahmed glaucoma valve (AGV) failure with a cloudy cornea. A 42-year-old patient with glaucoma underwent AGV implant surgery to treat secondary glaucoma due to chronic iridocyclitis in his left eye. After AGV, he developed bullous keratopathy (BK) in that eye. After Descemet stripping automated endothelial keratoplasty (DSAEK) was performed to treat BK, the intraocular pressure (IOP) increased and early failure of the DSAEK resulted again in a cloudy cornea. We could not precisely detect any cause of AGV failure with ordinary imaging instrumentation. An intraocular endoscope was used to determine that cause, and we found that the fibrous tissue occluded the tube of the AGV. The IOP decreased soon after the tissue was removed. We conclude that intraocular endoscopy was useful for diagnosing AGV failure with BK.