five

Auditory evoked potential P300 in adults: reference values

收藏
Figshare2016-06-01 更新2026-04-28 收录
下载链接:
https://figshare.com/articles/dataset/Auditory_evoked_potential_P300_in_adults_reference_values/20029513
下载链接
链接失效反馈
官方服务:
资源简介:
ABSTRACT Objective To establish reference intervals for cognitive potential P300 latency using tone burst stimuli. Methods This study involved 28 participants aged between 18 and 59 years. P300 recordings were performed using a two-channel device (Masbe, Contronic). Electrode placement was as follows: Fpz (ground electrode), Cz (active electrode), M1 and M2 (reference electrodes). Intensity corresponded to 80 dB HL and frequent and rare stimulus frequencies to 1,000Hz and 2,000Hz, respectively. Stimuli were delivered binaurally. Results Mean age of participants was 35 years. Average P300 latency was 305ms. Conclusion Maximum acceptable P300 latency values of 362.5ms (305 + 2SD 28.75) were determined for adults aged 18 to 59 years using the protocol described.
创建时间:
2016-06-01
5,000+
优质数据集
54 个
任务类型
进入经典数据集
二维码
社区交流群

面向社区/商业的数据集话题

二维码
科研交流群

面向高校/科研机构的开源数据集话题

数据驱动未来

携手共赢发展

商业合作