Comparison of the Fluid Resuscitation Rate with and without External Pressure Using Two Intraosseous Infusion Systems for Adult Emergencies, the CITRIN (Comparison of InTRaosseous infusion systems in emergency medicINe)-Study

收藏
NIAID Data Ecosystem2026-03-09 收录
下载链接:
https://figshare.com/articles/dataset/_Comparison_of_the_Fluid_Resuscitation_Rate_with_and_without_External_Pressure_Using_Two_Intraosseous_Infusion_Systems_for_Adult_Emergencies_the_CITRIN_Comparison_of_InTRaosseous_infusion_systems_in_emergency_medicINe_Study_/1617004
下载链接
链接失效反馈
官方服务:
资源简介:
Introduction Intraosseous infusion is recommended if peripheral venous access fails for cardiopulmonary resuscitation or other medical emergencies. The aim of this study, using body donors, was to compare a semi-automatic (EZ-IO®) device at two insertion sites and a sternal intraosseous infusion device (FASTR™). Methods Twenty-seven medical students being inexperienced first-time users were randomized into three groups using EZ-IO and FASTR. The following data were evaluated: attempts required for successful placement, insertion time and flow rates with and without external pressure to the infusion. Results The first-pass insertion success of the EZ-IO tibia, EZ-IO humerus and FASTR was 91%, 77%, and 95%, respectively. Insertion times (MW±SD) did not show significant differences with 17±7 (EZ-IO tibia) vs. 29±42 (EZ-IO humerus) vs. 33±21 (FASTR), respectively. One-minute flow rates using external pressures between 0 mmHg and 300 mmHg ranged between 27±5 to 69±54 ml/min (EZ-IO tibia), 16±3 to 60±44 ml/min (EZ-IO humerus) and 53±2 to 112±47 ml/min (FASTR), respectively. Concerning pressure-related increases in flow rates, negligible correlations were found for the EZ-IO tibia in all time frames (c = 0.107–0.366; p≤0.013), moderate positive correlations were found for the EZ-IO humerus after 5 minutes (c = 0.489; p = 0.021) and strong positive correlations were found for the FASTR in all time frames (c = 0.63–0.80; p≤0.007). Post-hoc statistical power was 0.62 with the given sample size. Conclusions The experiments with first-time users applying EZ-IO and FASTR in body donors indicate that both devices may be effective intraosseous infusion devices, likely suitable for fluid resuscitation using a pressure bag. Variations in flow rate may limit their reliability. Larger sample sizes will prospectively be required to substantiate our findings.
创建时间:
2015-12-02
二维码
社区交流群
二维码
科研交流群
商业服务