Limitation of diagnostic value of cone-beam CT in detecting apical root isthmuses
收藏Figshare2020-04-01 更新2026-04-28 收录
下载链接:
https://figshare.com/articles/dataset/Limitation_of_diagnostic_value_of_cone-beam_CT_in_detecting_apical_root_isthmuses/12056439
下载链接
链接失效反馈官方服务:
资源简介:
Abstract Isthmuses are reported as common anatomic complexities in teeth often associated with failures in endodontic treatment. They should be considered before starting treatment and a preoperative computed tomography scan (CT) may demonstrate these complexities. Objective To assess the diagnostic value of the highest resolution settings of a cone-beam CT (CBCT) system in identifying and measuring apical isthmuses, using micro-CT as reference. Methodology After micro-CT scanning, 40 humans’ lower first molars with isthmuses in the apical-3 mm of mesial roots were scanned by the highest resolution settings of the New Generation i-Cat ® CBCT equipment. Two blinded observers recorded the detection of isthmuses in CBCT scans. The lengths of isthmuses were compared between micro-CT and CBCT to assess the diagnostic value of CBCT. Quantitative data for sensitivity were represented as percentages (95% confidence interval). The Bland-Altman method was used to assess differences between gold standard lengths (micro-CT) and CBCT lengths. Results BCT demonstrated 30 positive findings, representing sensitivity for isthmus identification of 75% (95% CI=0.4114–1.1364). Differences between the lengths in micro-CT (1.99±0.40 mm) and CBCT (1.53±0.41 mm) were significant (p
创建时间:
2020-04-01



