Limitations of Stroke Volume Estimation by Non-Invasive Blood Pressure Monitoring in Hypergravity
收藏Figshare2016-01-15 更新2026-04-29 收录
下载链接:
https://figshare.com/articles/dataset/_Limitations_of_Stroke_Volume_Estimation_by_Non_Invasive_Blood_Pressure_Monitoring_in_Hypergravity_/1351099
下载链接
链接失效反馈官方服务:
资源简介:
BackgroundAltitude and gravity changes during aeromedical evacuations induce exacerbated cardiovascular responses in unstable patients. Non-invasive cardiac output monitoring is difficult to perform in this environment with limited access to the patient. We evaluated the feasibility and accuracy of stroke volume estimation by finger photoplethysmography (SVp) in hypergravity.MethodsFinger arterial blood pressure (ABP) waveforms were recorded continuously in ten healthy subjects before, during and after exposure to +Gz accelerations in a human centrifuge. The protocol consisted of a 2-min and 8-min exposure up to +4 Gz. SVp was computed from ABP using Liljestrand, systolic area, and Windkessel algorithms, and compared with reference values measured by echocardiography (SVe) before and after the centrifuge runs.ResultsThe ABP signal could be used in 83.3% of cases. After calibration with echocardiography, SVp changes did not differ from SVe and values were linearly correlated (pConclusionsNon-invasive ABP photoplethysmographic monitoring is an interesting technique to estimate relative stroke volume changes in moderate and sustained hypergravity. This method may aid physicians for aeronautic patient monitoring.
创建时间:
2016-01-15