five

Supplementary Material for: Feasibility of a Novel ECG Electrode Placement Method in Newborn Infants

收藏
Figshare2022-02-07 更新2026-04-28 收录
下载链接:
https://figshare.com/articles/dataset/Supplementary_Material_for_Feasibility_of_a_Novel_ECG_Electrode_Placement_Method_in_Newborn_Infants/19130441
下载链接
链接失效反馈
官方服务:
资源简介:
Background: International newborn resuscitation guidelines recommend electrocardiogram (ECG) heart rate (HR) monitoring at birth. We evaluated the application time of pre-set ECG electrodes fixed to a polyethene patch allowing adhesive-free attachment to the wet skin of the newborn chest. Objectives: Using a three-electrode pre-set ECG patch configuration, application success was calculated using video analysis and measured at three time points, the time to (1) apply electrodes; (2) detect recognizable QRS complexes after application; and (3) display a HR after application. Method: A prospective observational study in two UK tertiary maternity units was undertaken with 71 newborns including 23 who required resuscitation. Results: The median (IQR) time for ECG patch application was 8 (6–10) seconds, detection of recognizable QRS complexes 8 (2–12) seconds, and time to output HR was 23 (15–37) seconds. Conclusion: Pre-set ECG chest electrodes allow rapid HR information at birth without electrode detachment or compromising skin integrity.
创建时间:
2022-02-07
二维码
社区交流群
二维码
科研交流群
商业服务