Magnetic field inhomogeneities due to CO 2 incubator shelves: a source of experimental confounding and variability?
收藏DataONE2020-06-30 更新2025-04-19 收录
下载链接:
https://search.dataone.org/view/sha256:4fd0132ef91570af4edfa5cd3f9c7a8f5d1702480900bac19160a2bbdb5c65b4
下载链接
链接失效反馈官方服务:
资源简介:
A thorough assessment of the static magnetic field (SMF) inside a CO2 incubator allowed us to identify non-negligible inhomogeneities close to the floor, ceiling, walls and the door. Given that incubator's shelves are made of a non-magnetic stainless steel alloy, we did not expect any important effect of them on the SMF. Surprisingly, we did find relatively strong distortion of the SMF due to shelves. Indeed, our high-resolution maps of the SMF revealed that distortion is such that field intensities differing by a factor of up to 36 were measured on the surface of the shelf at locations only few millimetres apart from each other. Furthermore, the most intense of these fields was around five times greater than the ones found inside the incubator (without the metallic shelves in), while the lowest one was around 10 times lower, reaching the so-called hypomagnetic field range. Our findings, together with a survey of the literature on biological effects of hypomagnetic fields, soundly suppo...