Rapid assessment of lipidomics sample quality and quantity using attenuated total reflectance Fourier-transform infrared spectroscopy

收藏
NIAID Data Ecosystem2026-05-02 收录
下载链接:
https://zenodo.org/record/6592242
下载链接
链接失效反馈
官方服务:
资源简介:
In this work, we aimed to develop a simple lipid quality and quantification method for biological lipid extracts, as a step in lipidomics workflows, with minimal sample requirement. We chose FTIR spectroscopy with an Attenuate Total Reflectance (ATR) sampling method as it requires just 1 microliter of MS-ready sample without additional sample preparation. We validated the proposed lipidomics sample quality control workflow using a set of plasma samples (n=107, with 3-4 technical replicates) with comparison to LC-MS-based lipidomics. The following file contains the resulting spectra acquired by ATR-FTIR spectrometry for these plasma samples, standard curves and contaminated samples used for method development. Spectrometer was ambient blanked and detector cleaned between each measurement. Lipid samples were extracted by butanol-methanol (3:1) precipitation, and dried directly onto the ATR-FTIR detector. Absorbance was measured between 4,000 and 650 cm-1 wavenumbers, at a resolution of 8cm-1. Each spectra has been baseline corrected (whole spectra).