Insertion torque versus mechanical resistance of mini- implants inserted in different cortical thickness
收藏Figshare2014-06-01 更新2026-04-28 收录
下载链接:
https://figshare.com/articles/dataset/Insertion_torque_versus_mechanical_resistance_of_mini-_implants_inserted_in_different_cortical_thickness/20043592
下载链接
链接失效反馈官方服务:
资源简介:
OBJECTIVE: This study aimed to measure insertion torque, tip mechanical resistance to fracture and transmucosal neck of mini-implants (MI) (Conexão Sistemas de PróteseT), as well as to analyze surface morphology. METHODS: Mechanical tests were carried out to measure the insertion torque of MIs in different cortical thicknesses, and tip mechanical resistance to fracture as well as transmucosal neck of MIs. Surface morphology was assessed by scanning electron microscopy (SEM) before and after the mechanical tests. RESULTS: Values of mechanical resistance to fracture (22.14 N.cm and 54.95 N.cm) were higher and statistically different (P 0.05) to torsional fracture in the tip of MI (22.14 N.cm) when 3 mm cortical thickness (16.11 N.cm) and dense bone (23.95 N.cm) were used. Torsional fracture of the transmucosal neck (54.95 N.cm) was higher and statistically different (P
创建时间:
2014-06-01