A) A coronal T1w in and opposed phase Volumetric Interpolated Breath-hold Examination (VIBE) for calculation of the Dixon-based fat-/water-images as well as the μ-map. B) An axial, pre-contrast T1w Fast Low Angle Shot (FLASH). C) An axial T2w 2D half Fourier acquisition single shot turbo spin echo (HASTE). D) An axial diffusion weighted echo planar sequence (EPI DWI) with B values of 0, 500 and 1000 s/mm2. E) An axial T2w fat saturated turbo spin echo (TSE) sequence in breath-hold. F) A dynamic axial T1w VIBE. Four repetitive scans were performed (pre-contrast, arterial phase [20s delay], portal venous phase [60 sec delay], venous phase [100 sec delay]) after i.v.-injection of 0.1 ml/kg body weight Gadobutrol (Gadovist®, Bayer Healthcare, Berlin, Germany). G) An axial post-contrast T1w FLASH. Acquisition parameters for the applied MR-sequences (A-F).