Assessment of the light intensity of otoscopes utilized in teaching hospitals

收藏
DataCite Commons2025-06-01 更新2024-07-29 收录
下载链接:
https://scielo.figshare.com/articles/dataset/Assessment_of_the_light_intensity_of_otoscopes_utilized_in_teaching_hospitals/20014414/1
下载链接
链接失效反馈
官方服务:
资源简介:
ABSTRACT INTRODUCTION: To attain an accurate otoscopic diagnosis, a functioning device with adequate capacity must be used. OBJECTIVE: Evaluate the light intensity of otoscopes, comparing it utilizing the batteries present at the moment of calibration and after new batteries were supplied. METHODS: Cross-sectional study of a historical cohort, which assessed 38 otoscopes, measuring the light intensity with the batteries present at the moment of assessment compared to the intensity with new batteries, as well as charge of the test batteries and the new batteries. RESULTS: The mean of the sum of new batteries' charge was 3.19 V, and of the test batteries was 2.70 V, representing a decrease of 18.02% in charge. The mean luminosity with the new batteries was 366.89 lumens, whereas in the test batteries it was 188.32 lumens, representing a decrease of 83.75% in the light intensity. Student's t-test was applied for data comparison, showing a statistical difference between the light intensity with the original batteries and the new batteries (p = 0.0001; CI = 0.95). CONCLUSION: There was a statistically significant difference between the proportions of light intensity in the otoscopes. A small variation in battery charge results in a great variation in light.