Within-day reliability statistics.

收藏
Figshare2024-11-21 更新2026-04-28 收录
下载链接:
https://figshare.com/articles/dataset/Within-day_reliability_statistics_/27881689
下载链接
链接失效反馈
官方服务:
资源简介:
Assessing hydration status outside of laboratories can be challenging given that most hydration measures are invasive, stationary, costly, or have questionable validity. This study investigated the within-day, test-retest reliability, and minimal detectable change (MDC) of the MX3 Hydration Testing System (MX3 Diagnostics), a relatively low cost, noninvasive, and portable method to measure saliva osmolality. Seventy-five adults (44 men, 31 women; 29.6±10.8 yr, 171.1±9.2 cm, 79.1±15.4 kg) presented two saliva samples approximately 3 to 5 minutes apart. Fluid intake was avoided for at least 5 minutes prior to sample collections. For each sample collection, a researcher used the MX3 to tap the tip of a test strip to saliva on a participant’s tongue. Intra-class correlation coefficients (ICCs) and MDC were calculated as measures of reliability. Means for the two measurements were 64.3±19.0 mOsm and 65.5±17.9 mOsm (mean difference of -1.2±13.0 mOsm; t = -0.806, p = 0.423). Further analysis with a two-one-sided test procedure revealed equivalence between the two SOSM measurements (p 20 mOsm) in saliva osmolality outside of laboratory constraints.