Electrochemistry Boot Camp CV Data
收藏官方服务:
资源简介:
Data were collected on 09/20/2012 The electrochemical cell consisted of: Pt/Ir working electrode Ag/AgCl reference electrode Pt wire counter electrode Scan rates of 20mV/s, 50 mV/s, 100mV/s, 150 mV/s, and 200mV/s were tested A 10 sec quiet time was used for all scans (stir bar off after mixing for 30 sec) The solution was composed of: potassium ferrocyanide (4 mM) as redox probe potassium nitrate (1 mM) as electrolyte The testing conditions were: Temp = 21C (room temperature) pH=7.10 A BASi Epsilon potentiostat was used for all experiments Details and further analysis are shown in Vanegas et al (2015) link: https://pubs.rsc.org/en/content/articlelanding/2014/an/c4an01788a/unauth
提供机构:
Zenodo创建时间:
2024-05-11