Monte Carlo simulation for Compton suppression spectrometer
收藏Mendeley Data2023-02-23 更新2024-06-26 收录
下载链接:
https://data.mendeley.com/datasets/6s5ffj3j9w
下载链接
链接失效反馈官方服务:
资源简介:
Abstract
A Monte Carlo simulation program is presented that enables one to find the optimal position of the HPGe (high-purity germanium) detector and to design various geometrical topologies for the shield and detector. The program has been tested up to 2.8 MeV and been compared very well with the experimental results [B. Lal and K.V.K. Iyenger, Nucl. Instrum. Methods 79 (1970) 19].
Title of program: CSSM
Catalogue Id: ACVH_v1_0
Nature of problem
Due to the Compton scattering, the gamma-ray spectrum will be ruined by the Compton backgrounds, any small peak in the Compton continuum generated by the intensed gamma peak will become invisible. To overcome this problem, a Compton suppression spectrometer is designed for the reduction of the Compton backgrounds.
Versions of this program held in the CPC repository in Mendeley Data
ACVH_v1_0; CSSM; 10.1016/0010-4655(94)90033-7
This program has been imported from the CPC Program Library held at Queen's University Belfast (1969-2019)
创建时间:
2020-01-02