Non-invasive Instrument-Based Tests for Quantifying Anterior Chamber Flare in Uveitis: A Systematic Review

收藏
Figshare2020-04-07 更新2026-04-28 收录
下载链接:
https://figshare.com/articles/dataset/Non-invasive_Instrument-Based_Tests_for_Quantifying_Anterior_Chamber_Flare_in_Uveitis_A_Systematic_Review/12091557
下载链接
链接失效反馈
官方服务:
资源简介:
Purpose: Anterior chamber (AC) flare is a key sign for anterior uveitis. New instrument-based techniques for measuring AC flare can offer automation and objectivity. This review aims to identify objective instrument-based measures for AC flare. Methods: In this systematic review, we identified studies reporting correlation between instrument-based tests versus clinician AC flare grading, and/or aqueous protein concentration, as well as test reliability. Results: Four index tests were identified in 11 studies: laser-flare photometry (LFP), optical coherence tomography, ocular flare analysis meter (OFAM) and the double-pass technique. The correlation between LFP and clinician grading was 0.40–0.93 and 0.87–0.94 for LFP and protein concentration. The double-pass technique showed no correlation with clinician grading and insufficient information was available for OFAM. Conclusion: LFP shows moderate to strong correlation with clinician grading and aqueous protein concentration. LFP could be a superior reference test compared to clinician AC flare grading for validating new index tests.