five

Assessment of light transmission aggregometry on the routine coagulation analyzer Sysmex CS-2500 using CE-marked agonists from Hyphen Biomed

收藏
Figshare2019-04-08 更新2026-04-29 收录
下载链接:
https://figshare.com/articles/dataset/Assessment_of_light_transmission_aggregometry_on_the_routine_coagulation_analyzer_Sysmex_CS-2500_using_CE-marked_agonists_from_Hyphen_Biomed/7201340
下载链接
链接失效反馈
官方服务:
资源简介:
Light transmission aggregometry (LTA) is still considered as the “gold standard” for platelet function assessment but, as acompletely manual technology, it is labour intensive. This challenge can be overcome by performing platelet aggregometry in anautomated method on a routine coagulation analyzer. We aimed to compare and correlate results obtained from a traditional manual LTA solution realized in our Reference Center with an optimized automated system using CE-marked agonist reagents. Platelet rich plasma from patients with suspected platelet disorders, von Willebrand disease or antiplatelet therapy have been assessed using a wide range of agonist concentrations. Results were expressed as Maximal Platelet Aggregation and correlation was analyzed using the Passing and Bablok regression test. Platelet aggregometry studies were performed in 49 samples. Maximal aggregation response with ADP (0.5-10 μM), collagen (2 mg/μL), ristocetin (1.2 mg/mL) and arachidonic acid (1 mM) agonists showed significant correlation between the two aggregometers (p
创建时间:
2019-04-08
二维码
社区交流群
二维码
科研交流群
商业服务