Stimulus-response time to alarms of the intra-aortic balloon pump: safe care practices

收藏
DataCite Commons2020-08-31 更新2024-07-25 收录
下载链接:
https://scielo.figshare.com/articles/Stimulus-response_time_to_alarms_of_the_intra-aortic_balloon_pump_safe_care_practices/5666272
下载链接
链接失效反馈
官方服务:
资源简介:
ABSTRACT Objective: To characterize the sound alarms of the Intra-Aortic Balloon Pump (IABP) during aortic counterpulsation therapy; to measure the stimulus-response time of the team to these; and to discuss the implications of increasing this time for patient safety from the alarm fatigue perspective. Method: This is an observational and descriptive study with quantitative and qualitative approach, case study type, carried out in a Cardiac Surgical Intensive Care Unit. Results: The most audible IABP alarm was the one of high priority increased-reduced diastolic blood pressure. The stimulus-response time was 33.9 seconds on average. Conclusion: Managing the alarms of these equipment is essential to minimize the occurrence of the alarm fatigue phenomenon and to offer a safer assistance to patients who rely on this technology.