Load cells must be chosen to fit the conditions where the instrument will be used. Although creep is unavoidable in most high precisions sensors, the effect can be accounted for by choosing a low creep load cell and including zero measures in the logged data for continuous zero verification and adjustment. Zero creep affects bias only and has no effect on the scale factor ensuring measurement stability during long term deployment.