Simultaneous determination of four antibiotics in human plasma by high-performance LC–MS/MS: application to therapeutic drug monitoring: supplementary materials
收藏DataCite Commons2024-05-16 更新2025-04-15 收录
下载链接:
https://tandf.figshare.com/articles/dataset/Simultaneous_determination_of_four_antibiotics_in_human_plasma_by_high-performance_LC_MS_MS_application_to_therapeutic_drug_monitoring_supplementary_materials/22674388/1
下载链接
链接失效反馈官方服务:
资源简介:
Background: Aimed to simultaneously measure linezolid, voriconazole, cefoperazone and fluconazole
in human plasma suitable for therapeutic drug monitoring applications, a robust, rapid and easy-touse
HPLC–MS/MS approach was developed and validated. Materials & methods: Protein precipitation
was used to prepare analytes from 100 μl plasma. HPLC was employed for analyte separation, and
quantification was conducted via multiple reaction monitoring in positive ion mode. The methodology
was fully validated. Results & conclusion: All four antibiotics were found to be stable under the tested
conditions, and accuracy values ranged from 90.96 to 113.25% and CV values were <14.0%. This HPLC–
MS/MS method can be used for routine clinical therapeutic drug monitoring of linezolid, voriconazole,
cefoperazone and fluconazole simultaneously.
提供机构:
Taylor & Francis
创建时间:
2023-04-21