Investigation of the effect of cochlear implant electrode length on speech comprehension in quiet and noise compared with the results with users of electro-acoustic-stimulation, a retrospective analysis
收藏DataONE2020-06-24 更新2025-04-19 收录
下载链接:
https://search.dataone.org/view/sha256:df8b239fefd6a15880ef2d41d404656d4f52b25e8238149e88c9227d4bad6b29
下载链接
链接失效反馈官方服务:
资源简介:
Objectives: This investigation evaluated the effect of cochlear implant (CI) electrode length on speech comprehension in quiet and noise and compare the results with those of EAS users.
Methods: 91 adults with some degree of residual hearing were implanted with a FLEX20, FLEX24, or FLEX28 electrode. Some subjects were postoperative electric-acoustic-stimulation (EAS) users; the other subjects were in the groups of electric stimulation-only (ES-only).
Speech perception was tested in quiet and noise at 3 and 6 months of ES or EAS use. Speech comprehension results were analyzed and correlated to electrode length.
Results: While the FLEX20 ES and FLEX24 ES groups were still in their learning phase between the 3 to 6 months interval, the FLEX28 ES group was already reaching a performance plateau at the three months appointment yielding remarkably high test scores. EAS subjects using FLEX20 or FLEX24 electrodes outscored ES-only subjects with the same short electrodes on all 3 tests ...