Data for Bell-jar performance evaluation report
收藏NIAID Data Ecosystem2026-05-01 收录
下载链接:
https://zenodo.org/record/10219244
下载链接
链接失效反馈官方服务:
资源简介:
During the SI-Hg performance evaluation of elemental mercury gas generators on the market three generators were tested, e.g., PSA 10.536 elemental Hg generator, bell-jar and Tekran Model 3425. Key characteristics were determined e.g.; the stabilisation period, short-term drift, precision, i.e., reproducibility and repeatability of the concentration generated, linearity, bias, sensitivity to sample gas pressure, sensitivity to surrounding temperature and sensitivity to electrical voltage. All three generators could be tested according to the calibration protocol developed within the project. The results obtained with the different gas generator clearly shows the importance of a metrological calibration. All three candidate generators show a different bias for the setpoint compared to the calibrated output.
The data obtained during the performance evaluation of the bell-jar is published in this repository. The files of the following experiments can be found here:
m1
Measurement data online mercury analyser comparison VSL and bell-jar 2023-02-20
Bell_jar_calibration_m1
m2
Measurement data online mercury analyser comparison VSL and bell-jar 2023-02-28
Bell_jar_calibration_m2
m3
Measurement data online mercury analyser comparison VSL and bell-jar 2023-03-09
Bell_jar_calibration_m3
short-term drift
m1
Calibration bell-jar short term drif_m1
Bell_jar_drift_m1
m2
Calibration bell-jar short term drif_m2
Bell_jar_drift_m2
m3
Calibration bell-jar short term drif_m3
Bell_jar_drift_m3
m4
Calibration bell-jar short term drif_m4
Bell_jar_drift_m4
stability
Measurement data online mercury analyser comparison VSL and bell-jar stability
创建时间:
2023-11-29