Data for Identification algorithm of low-count energy spectra under short-duration measurement based on heterogeneous sample transfer
收藏DataCite Commons2025-04-27 更新2025-05-18 收录
下载链接:
https://www.scidb.cn/detail?dataSetId=0493c094da274f20aca099d40d6ed6a2
下载链接
链接失效反馈官方服务:
资源简介:
Measurement experiments in this study employed the AT6104DM spectrometer by Atomtex. The detection unit, enclosed within a seismic-resistant and waterproof stainless steel container, utilized a NaI(TI) crystal measuring Ø63x63mm to record gamma radiation from controlled radionuclides, exhibiting a typical resolution of 7.5\% at 662keV (${ }^{137}\mathrm{Cs}$). Detection energy range of the spectrometer spanned from 70keV to 3MeV. Measurement experiments were conducted in a laboratory setting. A specific open area within the laboratory was chosen as a fixed point. A ruler was placed on the ground, with the radiation source positioned at the origin of the ruler. The spectrometer was then moved to alter variables such as the relative distance from the source and the measurement angle. Two V-type radiation sources, ${ }^{137} \mathrm{Cs}$ and ${ }^{60}\mathrm{Co}$, as well as one IV-type radiation source, ${ }^{152}\mathrm{Eu}$, were utilized in the measurements, with activities of $3.7\times10^5$ Bq, $3.7\times10^5$ Bq, and $3.7\times10^8$ Bq, respectively. To investigate the proposed method's accuracy in identifying isotopes under shorter measurement times, the measurement duration was limited to 1 second to 5 seconds.
提供机构:
Science Data Bank
创建时间:
2024-09-30