PTR-ToF-MS data
收藏Figshare2018-12-18 更新2026-04-08 收录
下载链接:
https://figshare.com/articles/PTR-ToF-MS_data/7467746/1
下载链接
链接失效反馈官方服务:
资源简介:
Volatiles were analyzed with a PTR-ToF-MS 8000 (IoniconAnalytik GmbH, Innsbruck, Austria) using H<sub>3</sub>O<sup>+</sup> as reagent ion for the proton transfer reaction. The reaction takes place between H<sub>3</sub>O<sup>+</sup> ions and all the biogenic or anthropogenic VOCs having a proton affinity higher than that of water (165.2 kcal mol<sup>−1</sup>). Separation of single ions happens accordingly to their mass to charge (<i>m/z</i>) ratio. Drift applied voltage was set at 600 V, temperature at 110 °C, and pressure at 2.25 mbar. For each sample, about 6 grams of material was transferred in a glass jar provided with a special lid that allowed Teflon connection to a zero-air generator (inlet) and the PTR-ToF system (outlet). The head space was then measured by direct injection into the PTR-ToF drift tube inlet for 150 seconds, after respectively 1 and 20 min of exposure. Preliminary measurements on an empty jar were run before every experiment and used for background subtraction. All mass spectra up to <i>m/z</i> = 315 were simultaneously detected and recorded with 1 s as integration time. Internal calibration was based on <i>m</i>/<i>z</i> = 18.0338 (NH<sub>4</sub><sup>+</sup>), <i>m/z</i> = 21.0202 (H<sub>3</sub><sup>18</sup>O<sup>+</sup>) and <i>m/z</i> = 29.9974 (NO<sup>+</sup>).
创建时间:
2018-12-14