Dataset collected at BioMAX beamline, MAX IV Labortory, Sweden, of thaumatin. Data collection parameters:
Data collected: 30 November 2022
Wavelength: 0.9763 Å (12.7 keV)
Beamline: BioMAX, MAX IV Laboratory
Detector: DECTRIS EIGER 16M
File format: HDF5 in EIGER Nexus format
Number of frames: 1800
Oscillation per frame: 0.1°, 180° total
Resolution at detector corners: 2 Å
Estimated flux: 1.68e11 ph/s
Beamline parameters:
Synchrotron Name
MAX IV
Focusing Optics
K-B Mirror
Detector Type
Hybrid Pixel Counting Detector
Beamline name
BioMAX
Monochromator
Type
Si(111)
Detector Model
EIGER 16M
Synchrotron Current
297.0 mA
Beam Shape
ellipse
Detector Manufacturer
DECTRIS
Undulator type
Beam size at sample Horiz/vert (µm)
50/50
Detector pixel size (um)
75
Undulator gap
5.04 mm
Beam divergence
Horiz/vert (µrad)
6/104
Beam polarization
0.99
Crystals of thaumatin were grown from solutions of thaumatin (20-50 mg/mL) in NaK-tartrate (1.0 M), HEPES pH 7 (100 µM), and 25% glycerol.