five

Use of Weighted Performance Goals in Prospective Single-Arm Clinical Studies Designed to Assess the Safety and Effectiveness of Medical Devices

收藏
NIAID Data Ecosystem2026-03-11 收录
下载链接:
https://figshare.com/articles/dataset/Use_of_Weighted_Performance_Goals_in_Prospective_Single-Arm_Clinical_Studies_Designed_to_Assess_the_Safety_and_Effectiveness_of_Medical_Devices/12844374
下载链接
链接失效反馈
官方服务:
资源简介:
In addition to extensive preclinical testing, commercialization of medical devices in the United States under some circumstances may be supported with clinical data derived from single arm studies where the clinical endpoint is compared with a performance goal (PG). Since the performance is expected to be different between two clinically defined mutually exclusive patient subpopulations, the use of weighted PG is often adopted. In this article, we discuss the limitations of the current practice where weights need to be prespecified and fixed, and we propose an alternative approach that relaxes the fixed weight requirements.
创建时间:
2020-08-21
二维码
社区交流群
二维码
科研交流群
商业服务