Abstract This study developed a method for furosemide quantification through high performance liquid chromatographic technique. Special attention was given to solute loss and storage stability due to furosemide’s low solubility and photosensitivity, respectively. The performance of Nylon and PVDF filters was tested in a 2 mg.L-1 furosemide solution. PVDF filters showed better recovery capacity and therefore are more suitable for furosemide filtration. Over eight days, three different storage conditions were studied to access furosemide degradation susceptibility: (i) exposure to light at room temperature, (ii) storage at room temperature without exposure to light, and (iii) storage at 4ºC without exposure to light. The study demonstrated that after 48 h under natural light exposure furosemide was completely degraded. Furosemide solution stored in the dark was stable. Storage temperature did not seem to affect furosemide concentration. The study shows that the selection of more suitable filter and storage conditions for furosemide determination is crucial to avoid underestimation errors.