Closed-loop control of k-space sampling via physiologic feedback for cine MRI
收藏DataONE2020-12-15 更新2025-05-10 收录
下载链接:
https://search.dataone.org/view/sha256:32971df82830e3fc314ea8aefbdb39ae4801f7b2ed7e8dbaeffe3acc84ba7595
下载链接
链接失效反馈官方服务:
资源简介:
This dataset accompanies the manuscript outlining a method for closed-loop sampling of k-space in response to physiologic changes. The closed-loop approach enables near-uniform radial sampling in a segmented acquisition approach which was higher than predetermined golden-angle radial sampling. This can be utilized to increase the sampling or decrease the temporal footprint of an acquisition and the closed-loop framework has the potential to be applied to patients with complex heart rhythms.
Briefly, Segmented cine cardiac MRI combines data from multiple heartbeats to achieve high spatiotemporal resolution cardiac images, yet predefined k-space segmentation trajectories can lead to suboptimal k-space sampling. In this work, we developed and evaluated an autonomous and closed-loop control system for radial k-space sampling to increase sampling uniformity.
The dataset includes both the algorithm and the data used in our manuscript. Our closed-loop system autonomously selects radial k...
创建时间:
2025-04-28