five

Investigation of X rays beams uniformity in radiotherapeutic tumor treatment procedure using LuAG:Ce crystal detectors

收藏
DataCite Commons2025-03-18 更新2025-04-16 收录
下载链接:
https://repod.icm.edu.pl/citation?persistentId=doi:10.18150/QTAFPS
下载链接
链接失效反馈
官方服务:
资源简介:
In this study, Ce3+ doped Lu3Al5O12 garnet (LuAG) crystal detectors, with a density of 6.0 g/cm³ and an effective atomic number (Zeff) of 62, are proposed as promising materials for radiotherapy applications. This type of detector demonstrate excellent uniformity of structural and optical properties, high thermoluminescence (TL) light yield, optimal position of main TL glow peaks at temperatures around 245–295°C, and high radiation stability. The set of TL detectors made from LuAG single crystal were used to evaluate the uniformity of dose and energy spectra of X-ray radiation from a clinical accelerator with 6 MV and 15 MV beams at the Department of Medical Physics, Oncology Center in Bydgoszcz, Poland, and γ-rays from a 60Co source at the National Institute of Oncology in Warsaw. The LuAG crystal detectors demonstrated highly promising results for registering X-ray radiation from accelerators with both 6 MV and 15 MV electron beams, as well as γ-rays from a 60Co source with energies of 1.17 and 1.33 MeV.