CT simulation of MR-Linac patients without the use of a CT table overlay
收藏DataCite Commons2023-12-08 更新2024-08-18 收录
下载链接:
https://figshare.com/articles/dataset/CT_simulation_of_MR-Linac_patients_without_the_use_of_a_CT_table_overlay/24777396
下载链接
链接失效反馈官方服务:
资源简介:
While current MR-Linac (MRL) treatment workflows utilize a large table overlay during CT simulation to convert indexing between the two machines, we developed a look-up-table (LUT) as an alternative approach. After populating the LUT, index conversion factors were verified at three separate table locations. The resultant root-mean-square isocenter shifts on the MRL were 0.04/0.08 cm, 0.08/0.07 cm, and 0.09/0.08 cm with/without using the table overlay during simulation in the lateral, longitudinal, and vertical directions, respectively, which is within registration tolerance. Clinical implementation of the LUT has resulted in a more efficient MRL treatment workflow while maintaining accurate patient setup.
提供机构:
figshare
创建时间:
2023-12-08



