CIMAC CyTOF Multi-Site Harmonization Vericells tSNE Files
收藏NIAID Data Ecosystem2026-03-12 收录
下载链接:
http://flowrepository.org/id/FR-FCM-Z342
下载链接
链接失效反馈官方服务:
资源简介:
The Cancer Immune Monitoring and Analysis Centers - Cancer Immunologic Data Commons (CIMAC-CIDC) initiative is to identify biomarkers of response to cancer immunotherapies, with CyTOF being one of the main platforms to do so. This experiment was conducted to evaluate data reproducibility and comparability on the CyTOF platform between different laboratories across the US when using a harmonized acquisition and staining protocol. This experiment contains the FCS files of the tSNE analysis done on the Vericells by gating on the Ta181+ cells in the "CIMAC CyTOF Multi-Site Harmonization Raw FCS" experiment (FlowRepository ID: FR-FCM-Z34Z)
Conclusion:
CyTOF data was reproducible across all CIMAC sites with coefficient of variances (CVs) under 20% for identifying percentage of major immune populations in PBMCs. Two sets of reference control samples were generated. 1) Pre-stained: 4 PBMC donor lots spiked with Ta181 labeled Veri-cells were stained at a single site and shipped to all CIMAC sites in order to evaluate instrument acquisition reproducibility. 2) Site-stained: 4 PBMC donor lots along with an identical lot of lyophilized antibody cocktails were shipped to each CIMAC site and stained at each CIMAC site with the spike in Veri-cells in order to evaluate staining reproducibility.