Reliability of single-use PEEP-valves attached to self-inflating bags during manual ventilation of neonates â an in vitro study
收藏资源简介:
Introduction International resuscitation guidelines suggest to use positive end-expiratory pressure (PEEP) during manual ventilation of neonates. Aim of our study was to test the reliability of self-inflating bags (SIB) with single-use PEEP valves regarding PEEP delivery and the effect of different peak inflation pressures (PIP) and ventilation rates (VR) on the delivered PEEP. Methods Ten new single-use PEEP valves from 5 manufacturers were tested by ventilating an intubated 1kg neonatal manikin containing a lung model with a SIB that was actuated by an electromechanical plunger device. Standard settings: PIP 20cmH2O, VR 60/min, flow 8L/min. PEEP settings of 5 and 10cmH2O were studied. A second test was conducted with settings of PIP 40cmH2O and VR 40/min. The delivered PEEP was measured by a respiratory function monitor (CO2SMO+). Results Valves from one manufacturer delivered no relevant PEEP and were excluded. The remaining valves showed a continuous decay of the delivered pres...