RAD, limits of agreement analysis, and Clarke error grid analysis between CGMS and CLD values in data sets at different time after CGMS sensor calibration and in different glycemic ranges.
收藏资源简介:
Comparisons of RAD and bias were performed by unpaired t-test. Difference in percentage of matched points in Clarke error grid zone A was analyzed by χ2 test. RAD was calculated as absolute difference between time-matched CGMS and CLD value divided by CLD value, multiplied by 100. Bias was defined as the mean of the difference between time-matched CGMS and CLD values (CGMS minus CLD). Euglycemic level was defined as CLD serum glucose concentration of 3.9 to 10.0 mmol/L, and hyperglycemic level was defined as CLD serum glucose concentration greater than 10.0 mmol/L. Upper and lower limit of agreement were defined as bias±1.96 SD of the mean bias. CGMS = continuous glucose monitoring system. CLD = central laboratory device. RAD = relative absolute difference.