HPLC method development and validation for the estimation of Axitinib in rabbit plasma
收藏DataCite Commons2022-06-09 更新2024-07-25 收录
下载链接:
https://scielo.figshare.com/articles/dataset/HPLC_method_development_and_validation_for_the_estimation_of_axitinibe_in_rabbit_plasma/5668204
下载链接
链接失效反馈官方服务:
资源简介:
ABSTRACT A rapid, sensitive, and accurate high performance liquid chromatography for the determination of Axitinib (AN) in rabbit plasma is developed using crizotinibe as an internal standard (IS). Axitinib is a tyrosine kinase inhibitor, used in the treatment of advanced kidney cancer, which works by slowing or stopping the growth of cancer cells. The chromatographic separation was performed on a Waters 2695, Kromosil (150 mm × 4.6 mm, 5 µm) column using a mobile phase containing buffer (pH 4.6) and acetonitrile in the ratio of 65:35 v/v with a flow rate of1 mL/min. The analyte and internal standard were extracted using liquid-liquid extraction with acetonitrile. The elution was detected by photo diode array detector at 320 nm.The total chromatographic runtime is 10.0 min with a retention time for Axitinib and IS of 5.685, and 3.606 min, respectively. The method was validated over a dynamic linear range of 0.002-0.2µg/mL for Axitinib with a correlation coefficient of r2 0.999.
提供机构:
SciELO journals
创建时间:
2017-12-05