five

Temperature correction for cylindrical cavity perturbation measurements

收藏
DataCite Commons2024-12-05 更新2024-07-13 收录
下载链接:
https://research-data.cardiff.ac.uk/articles/dataset/Temperature_correction_for_cylindrical_cavity_perturbation_measurements/27051382
下载链接
链接失效反馈
官方服务:
资源简介:
This dataset demonstrates a method for temperature correcting microwave cavity perturbation measurements by monitoring two modes; one which is perturbed by the sample and one which is not (referred to as a nodal mode). The nodal modes used (TM310 and TE311 for an axial sample in a cylindrical cavity) are subject only to sample-independent influences. To demonstrate this technique the bulk permittivity of a PTFE rod has been measured under varying temperature conditions. The results show that without correction, the measured temperature dependent dielectric constant has large variations associated with the stepped and linear temperature ramping procedures. The corrected response mitigates systematic errors in the real part. However, the correction of the imaginary part requires careful consideration of the mode coupling strength.
提供机构:
Cardiff University
创建时间:
2017-02-21
5,000+
优质数据集
54 个
任务类型
进入经典数据集
二维码
社区交流群

面向社区/商业的数据集话题

二维码
科研交流群

面向高校/科研机构的开源数据集话题

数据驱动未来

携手共赢发展

商业合作