Data from: Field test and validation of the DYNEX Technologies Inc. Multiplier® measles, mumps, rubella and varicella zoster multiplexed assay system in the Democratic Republic of the Congo using dried blood spots
收藏Mendeley Data2024-06-25 更新2024-06-27 收录
下载链接:
https://zenodo.org/records/4956860
下载链接
链接失效反馈官方服务:
资源简介:
Here we describe baseline validation studies and field performance of a research-use-only chemiluminescent multiplex serology panel for measles, mumps, rubella, and varicella-zoster virus used with dried blood spots in support of the 2013-2014 Democratic Republic of the Congo Demographic and Health Survey. Characterization of the panel using US-FDA-cleared commercial kits shows good concordance for measles, mumps, rubella and varicella zoster with average sensitivity across assays of 94.9% and an average specificity of 91.4%. As expected, performance versus available standards validated for plaque-reduction assays does not provide a 1:1 correspondence with International Units, yet demonstrates excellent linearity (average Hill's Slope = 1.02) and ~4 logs of dynamic range. In addition, for the four specific assays, the multiplexed format allowed for inclusion of three positive and two negative controls for each sample. A prototype DYNEX Multiplier® chemiluminescent automated immunoassay instrument with a Charged Coupled Device camera provided a rugged and robust processing and data acquisition platform. Performance of a multiplex instrument for serological testing in a substantially resource-limited environment shows excellent reproducibility, minimal cross-reactivity, and a clear discrimination between specific assays, and should be considered a viable option for future serosurveys.