Metabolomics analysis of multiple organs upon in vivo application of WMK-1
收藏Zenodo2025-12-24 更新2026-05-26 收录
下载链接:
https://zenodo.org/doi/10.5281/zenodo.17985398
下载链接
链接失效反馈官方服务:
资源简介:
C57BL/6 mice were treated with vehilce or 40 mpk WMK-1 for two days (QD, through i.p. injection). Tissues, including liver, spleen, kidney, lung, and heart were collected and loaded for lipidomics and metabolomics analysis via LC-MC/MS.
For tissue metabolomics profiling, 3 µL of metabolomics sample was loaded into a Vanquish Flex system coupled with an ACQUITY UPLC BEH Amide column (100 × 2.1 mm, 1.7 μm, 130 Å, Waters, Cat# 186004741), and separated chromatographically over a 15-min LC-gradient elution at a flow rate of 0.3 mL/min. The mobile phase for the positive ion acquisition mode contains H2O with 0.1% (v/v) formic acid and 10 mM ammonium formate for Phase A, and ACN/H2O (95:5) with 0.1% (v/v) formic acid and 10 mM ammonium formate for Phase B. The mobile phase for the negative ion acquisition mode is similar to the buffers for the positive ion mode except excluding formic acid in the buffer
The Compound Discoverer v3.2 software (Thermo Fisher Scientific) was used for LC-MS/MS metabolomics data processing including peak alignment, background filtering, peak calling, and compound identification. We applied an adaptive curve alignment model and filtered out metabolites with intensities lower than the minimum cutoff of 100,000 in both positive and negative ion mode datasets. The remaining metabolites were used for further analysis. Adduct ions including [M+H]+, [M+Na]+, [M+K]+, [M+NH4]+and [M-H]- were identified by assigning predicted chemical composition to each ion based on the exact mass within a ±5 ppm deviation and the isotopic pattern with intensity tolerance < 30%.
For metabolite annotation, mzCloud search was used for mapping the MS/MS spectra with the endogenous metabolite library. A cosine-based identity search procedure with match activation type was employed during the DDA search method, which enables accurate matching of fragmentation patterns. During the mapping of activation energy, a tolerance within 20, and a minimum intensity threshold for filtering low-intensity signals were employed. Similarity search was turned off during annotation. We used a match factor threshold of 60 during the metabolite identification. Metabolites not detected in the samples were substituted with background noise signals. A minimum of 50% QC coverage and a maximum QC area RSD of 30% were employed during the QC correction. Lastly, metabolites whose intensities are lower than five times that of the corresponding blank samples were filtered out.
提供机构:
Zenodo
创建时间:
2025-12-24



