Accuracy and user performance evaluation of a blood glucose monitoring system which wirelessly transmits results to compatible insulin pumps
收藏Figshare2020-02-25 更新2026-04-28 收录
下载链接:
https://figshare.com/articles/dataset/Accuracy_and_user_performance_evaluation_of_a_blood_glucose_monitoring_system_which_wirelessly_transmits_results_to_compatible_insulin_pumps/11897994
下载链接
链接失效反馈官方服务:
资源简介:
Objective: To perform evaluations of the CONTOUR PLUS LINK 2.4 blood glucose monitoring system (BGMS) assessed according to ISO 15197:2013 criteria. Methods: Clinical trial registered at ClinicalTrials.gov (NCT01824355). In a laboratory study (Study 1), capillary fingertip blood samples from 100 subjects were evaluated in duplicate, using three test strip lots. In a clinical trial (Study 2), 113 adults with diabetes were enrolled, and BGMS results and Yellow Springs Instruments (YSI) analyzer reference measurements were compared for subject- and trial staff-obtained fingertip blood and subject-obtained palm blood. Subjects completed a questionnaire to evaluate BGMS ease of use. Results: In Study 1, 100% of combined results (all test strip lots) fulfilled ISO 15197:2013 section 6.3 criteria. In Study 2, 97.7% of subject-obtained fingertip results and 100% of trial staff-obtained fingertip results met ISO 15197:2013 section 8 criteria. Additionally, 93.8% of palm results were within ± 15 mg/dL of mean YSI measurements for glucose concentrations Conclusion: The BGMS exceeded minimum ISO 15197:2013-specified accuracy criteria in the laboratory and in the hands of lay users with diabetes.