MOESM1 of Evaluation of automated malaria diagnosis using the Sysmex XN-30 analyser in a clinical setting
收藏资源简介:
Additional file 1: Fig. S1. XN-30 software improvements. a. In the prototype software, the M scattergram plot area was divided into 3 regions classifying signals as WBCs (cluster a), non-infected RBCs, platelets and debris (cluster b) and MI-RBCs (cluster c). In brief, the gating strategy sequence was to first identify cluster a and then cluster b. In the prototype, all remaining signals were then assigned to cluster c. Upon further investigation, it was identified that the specified MI-RBC area (cluster c) was too broad, giving rise to false positive MI-RBC results. b. The software was subsequently improved for the XN-30 by narrowing the MI-RBC area (cluster c) and incorporating the recognition of one or more distinct clusters (ring forms, gametocytes, trophozoites/schizonts) within an appropriate shape and position as a prerequisite for generating an MI-RBC result. Fig. S2. XN-30 malaria classification algorithm. The left side illustrates the algorithm flow if a malaria cluster is not recognized. In such cases, if the MI-RBC# is