Mass spectrometry (MS) is an attractive alternative to quantification of proteins by immunoassays, particularly for protein biomarkers of clinical relevance. Reliable quantification requires that the MS-based assays are robust, selective, and reproducible. Thus, the development of standardized protocols is essential to introduce MS into clinical research laboratories. The aim of this study was to establish a complete workflow for assessing the transferability and reproducibility of selected reaction monitoring (SRM) assays between clinical research laboratories. Four independent laboratories in North America, using identical triple-quadrupole mass spectrometers (Quantum Ultra, Thermo), were provided with standard protocols and instrumentation settings to analyze unknown samples and internal standards in a digested plasma matrix to quantify 51 peptides from 39 human proteins using a multiplexed SRM assay. The interlaboratory coefficient of variation (CV) was less than 10% for 25 of 39 peptides quantified (12 peptides were not quantified based upon hydrophobicity) and exhibited CVs less than 20% for the remaining peptides. In this report, we demonstrate that previously developed research platforms for SRM assays can be improved and optimized for deployment in clinical research environments.