Agreement of ZHF thermometry compared with tympanic temperature monitoringTB-D-23-00757
收藏NIAID Data Ecosystem2026-05-01 收录
下载链接:
https://data.mendeley.com/datasets/2vmxwc257w
下载链接
链接失效反馈官方服务:
资源简介:
This study aims to compare the application of the cutaneous zero-heat-flux (ZHF) thermometer in general anesthesia using the infrared tympanic measurement as a reference. We conducted a prospective observational study and enrolled in 130 patients scheduled for major surgery with general anesthesia. Core temperature was continuously measured using a forehead ZHF sensor (Tzhf) and an infrared tympanic thermometer (Ttym). Comparing the paired measurement of Tzhf and Ttym, we assessed the agreement using Bland-Altman analysis and concordance correlation coefficient (CCC). We further calculated the percentage of difference within 0.5°C between the two devices. Sensitivity, specificity, and predictive values were estimated for interpretation the performance of the ZHF thermometer in detecting hypothermia and hyperthermia.
创建时间:
2024-01-12