When the flat-bottomed, glass-stoppered, round bottle traditionally used to make VDRL antigen was discontinued, an appropriate substitute was needed. Although many laboratories have switched to one of the other nontreponemal tests for syphilis serology screening, the VDRL test remains the only approved procedure for testing spinal fluids of patients with possible neurosyphilis. We tested 25-ml glass-stoppered, convex-bottomed Erlenmeyer flasks to determine if these could be used as appropriate substitutes. We tested 52 reactive sera and 54 nonreactive sera by using one reference antigen prepared in the traditional flat-bottomed bottles and five antigens prepared in the Erlenmeyer flasks. Results with all serum samples were comparable. We also tested two lots of a commercial antigen plus an additional lot of reference antigen. Again there was no difference in the reactivity of the antigens. Therefore, we conclude that 25-ml glass-stoppered Erlenmeyer flasks can be used as an appropriate substitute for glass-stoppered, flat-bottomed, round glass bottles in the making of VDRL antigen.