Comparison of PBMC sample aliquots acquired in parallel on a CyTOF2 and a Helios mass cytometer
收藏NIAID Data Ecosystem2026-04-25 收录
下载链接:
http://flowrepository.org/id/FR-FCM-Z2ZJ
下载链接
链接失效反馈官方服务:
资源简介:
To evaluate the performance of Fluidigm's 3rd generation Helios mass cytometer compared to the 2nd generation CyTOF2 on PBMC sample aliquots.
Conclusion:
We find divergent behavior and performance between biological samples and the synthetic EQ beads used for instrument calibration and acquisition normalization. While the Helios mass cytometer resulted in higher median signal intensities and lower coefficients of variance (CV) on the synthetic EQ beads, we observed reduced median signal and higher CVs on the PMBC sample aliquots when compared to the CyTOF2. One stained PBMC sample was divided into multiple aliquots for acquisition on the same CyTOF2 and Helios instruments.