Acquisition parameters for the applied MR-sequences (A-F).
收藏Figshare2015-12-03 更新2026-04-29 收录
下载链接:
https://figshare.com/articles/dataset/_Acquisition_parameters_for_the_applied_MR_sequences_A_F_/1535154
下载链接
链接失效反馈官方服务:
资源简介:
A) A coronal T1w in and opposed phase Volumetric Interpolated Breath-hold Examination (VIBE) for calculation of the Dixon-based fat-/water-images as well as the μ-map.B) An axial, pre-contrast T1w Fast Low Angle Shot (FLASH).C) An axial T2w 2D half Fourier acquisition single shot turbo spin echo (HASTE).D) An axial diffusion weighted echo planar sequence (EPI DWI) with B values of 0, 500 and 1000 s/mm2.E) An axial T2w fat saturated turbo spin echo (TSE) sequence in breath-hold.F) A dynamic axial T1w VIBE. Four repetitive scans were performed (pre-contrast, arterial phase [20s delay], portal venous phase [60 sec delay], venous phase [100 sec delay]) after i.v.-injection of 0.1 ml/kg body weight Gadobutrol (Gadovist®, Bayer Healthcare, Berlin, Germany).G) An axial post-contrast T1w FLASH.Acquisition parameters for the applied MR-sequences (A-F).
创建时间:
2015-12-03