DNA ploidy analysis in B-ALL MRD using FxCycle violet
收藏NIAID Data Ecosystem2026-03-09 收录
下载链接:
http://flowrepository.org/id/FR-FCM-ZZM4
下载链接
链接失效反馈官方服务:
资源简介:
To demonstrate application of DNA ploidy analysis using Fxcycle violet (FCV) and simultaneous multicolor (4-7) surface immunophenotyping.
Conclusion:
This method DNA ploidy analysis using FxCycle violet and simultaneous six color immunophenotyping can be used flow cytometric DNA ploidy determination in B-ALL MRD evaluation (up to 100 events).
Notes:
This experiment proves that this method can be useful for DNA ploidy analysis for B-ALL MRD analysis. However, it is recommended acquire at least 200 events of tumor cells for good results for rare events of tumor. Although, CV of G1 peak might be slightly higher (between 4 to 6), it is acceptable as it is still falls within the 10% standard error and does not affect the results of the assay. Few unstained and apoptotic cells always present in the sample showing decreasing staining pattern of FxCycle. However, these does not interfere with analysis and interpretation once gated out. DNA ploidy and simultaneous six color immunophenotyping using FxCycle-Violet.
创建时间: