遇见数据集

Microfluidic Device for Patient-Centric Multiplexed Assays with Readout in Centralized Laboratories

收藏
NIAID Data Ecosystem2026-03-14 收录
官方服务:

资源简介:

Patient-centric sampling strategies, where the patient performs self-sampling and ships the sample to a centralized laboratory for readout, are on the verge of widespread adaptation. However, the key to a successful patient-centric workflow is user-friendliness, with few noncritical user interactions, and simple, ideally biohazard-free shipment. Here, we present a capillary-driven microfluidic device designed to perform the critical biomarker capturing step of a multiplexed immunoassay at the time of sample collection. On-chip sample drying enables biohazard-free shipment and allows us to make use of advanced analytics of specialized laboratories that offer the needed analytical sensitivity, reliability, and affordability. Using C-Reactive Protein, MCP1, S100B, IGFBP1, and IL6 as model blood biomarkers, we demonstrate the multiplexing capability and applicability of the device to a patient-centric workflow. The presented quantification of a biomarker panel opens up new possibilities for e-doctor and e-health applications.