Comparison of mass specra obtained with and Atom Probe installed in a TEM (JEOL F 200, designated as SATMET) at Groupe de Physique des Matériaux UMR 6634 (Saint Etienne du Rouvray, France) and in a LEAP 5000 XS
收藏官方服务:
资源简介:
Comparison of mass specra obtained with an Atom Probe installed in a TEM (JEOL F 200, designated as SATMET) at Groupe de Physique des Matériaux UMR 6634 (Saint Etienne du Rouvray, France) and in a LEAP 5000 XS Conditions of analyses Material: Fe-51.4Cr (at%) alloy Temperature of APT analyses: 78KPulse repetition rate in LEAP 5000 XS: 25 kHzPulse repetition rate in SATMET: 20 kHzNumber of events collected in LEAP 5000 XS: 807 784Number of events collected in SATMET: 472 423