Data for: REBEn 2024-0485: Comparison between two methods of cleaning and disinfection of digital pulse oximeters in primary care
收藏SciELO Data2024-10-23 更新2026-04-25 收录
下载链接:
https://data.scielo.org/dataset.xhtml?persistentId=doi:10.48331/SCIELODATA.G7FPZG
下载链接
链接失效反馈官方服务:
资源简介:
Data is in CSV format and express results of Microbiological culture (expressed in colony forming units) and adenosine triphosphate (ATP) assay (expressed in relative light units) results obtained from oximeters inoculated with Salmonella typhimurium, Staphylococcus aureus or Escherichia coli: after water and detergent followed by 70% ethanol (alcohol); after one-step disinfectant based on benzalkonium chloride and polyhexamethylene biguanide hydrochloride (PHMB); and no cleaning and disinfection (control). To evaluate the duration of both methods, we cleaned and disinfected 60 devices (30 per technique), in batches of 10 oximeters, to ensure routine processing conditions. We registered the duration of the procedure from the beginning of cleaning and disinfection of the first device until the end of the process of the last device of each batch. These data is also included in the CSV archive - duration expressed in minutes. (2024-10-10)
创建时间:
2024-10-23