five

Accuracy of speckle tracking in the context of stress echocardiography in short axis view: An in vitro validation study

收藏
Figshare2018-03-28 更新2026-04-29 收录
下载链接:
https://figshare.com/articles/dataset/Accuracy_of_speckle_tracking_in_the_context_of_stress_echocardiography_in_short_axis_view_An_in_vitro_validation_study/6042617
下载链接
链接失效反馈
官方服务:
资源简介:
AimThis study aimed to test the accuracy of a speckle tracking algorithm to assess myocardial deformation in a large range of heart rates and strain magnitudes compared to sonomicrometry.Methods and resultsUsing a tissue-mimicking phantom with cyclic radial deformation, radial strain derived from speckle tracking (RS-SpT) of the upper segment was assessed in short axis view by conventional echocardiography (Vivid q, GE) and post-processed with clinical software (EchoPAC, GE). RS-SpT was compared with radial strain measured simultaneously by sonomicrometers (RS-SN). Radial strain was assessed with increasing deformation rates (60 to 160 beats/min) and increasing pulsed volumes (50 to 100 ml/beat) to simulate physiological changes occurring during stress echocardiography. There was a significant correlation (R2 = 0.978, P ConclusionsSpeckle tracking demonstrated a good correlation with sonomicrometry for the assessment of radial strain independently of the heart rate and strain magnitude in a physiological range of values. Though speckle tracking seems to be a reliable and reproducible technique to assess myocardial deformation variations during stress echocardiography, further studies are mandated to analyze the impact of angulated and artefactual out-of-plane motions and inter-vendor variability.
创建时间:
2018-03-28
二维码
社区交流群
二维码
科研交流群
商业服务