Calculated biases between SaO2 (oxyhemoglobin saturation measured by blood gas analyzer [ABL 825]) and SpO2 (oxyhemoglobin saturation measured by 3 pulse oximeters) among patients under room air condition.
收藏数据链接:
官方服务:
资源简介:
The calculated mean biases (SpO2—SaO2) measured by 3 pulse oximeters for the 85%–100% saturation range are shown. Data are presented as the mean ± SD and analyzed by one-way analysis of variance followed by Tukey’s test.*pCalculated biases between SaO2 (oxyhemoglobin saturation measured by blood gas analyzer [ABL 825]) and SpO2 (oxyhemoglobin saturation measured by 3 pulse oximeters) among patients under room air condition.
创建时间:
2015-12-03