Four sweep cubic insulin data set recorded using dose mode (1MGy / scan) on Diamond Light Source beamline i04
收藏NIAID Data Ecosystem2026-05-01 收录
下载链接:
https://zenodo.org/record/8376817
下载链接
链接失效反馈官方服务:
资源简介:
Four sweep cubic insulin data set recorded on Diamond Light Source beamline i04 as part of routine testing. Data were collected with four full 360° scans / 0.1° / image to give a total of 1MGy / scan i.e. "carefully" to avoid problems of obvious radiation damage. The beam size was adjusted to be about the same as the crystal, spreading the dose uniformly across the sample.
Processing using my "standard DIALS script" gives nice data to 1.2Å:
Overall Low High
High resolution limit 1.20 3.26 1.20
Low resolution limit 55.07 55.13 1.22
Completeness 100.0 100.0 100.0
Multiplicity 150.2 155.4 76.0
I/sigma 40.8 295.5 0.4
Rmerge(I) 0.075 0.027 7.170
Rmerge(I+/-) 0.075 0.026 7.127
Rmeas(I) 0.076 0.027 7.218
Rmeas(I+/-) 0.075 0.027 7.220
Rpim(I) 0.006 0.002 0.815
Rpim(I+/-) 0.008 0.003 1.140
CC half 1.000 1.000 0.335
Anomalous completeness 100.0 100.0 100.0
Anomalous multiplicity 77.7 84.8 38.9
Anomalous correlation 0.499 0.680 -0.005
Anomalous slope 0.626
dF/F 0.031
dI/s(dI) 0.761
Total observations 3709553 203869 91461
Total unique 24699 1312 1204
Data shared to allow use in tutorials etc. as well as a reference for future data handling questions.